Re: NMRView peak-picking



Hmmmm, I've got a theory about why this is happening. I'll check it out. Hmmm, now I've checked it out. There is a problem based on the relative size (as expressed in datapoints) of the window in the x, y and z dimensions. I'll fix it this weekend.


Delete Button : should delete any peaks in the region enclosed by the cursor box. Restore Button: should undelete them, if the peak list hasn't already been compressed.

I'll check to make sure these are still working.
On May 18, 2007, at 2:40 PM, David A. Horita wrote:

Hi,
I'm seeing perplexing behavior when peakpicking (v 7.0.15); If I pick a 3D spectrum in "replace" mode, I get peaks picked in the wrong places. This does not happen if I set Region=Window and am displaying the full spectrum. If I set Region=Box, the peaks are offset. If I've expanded the spectrum, this happens whether Region=Box or Window. If I set Mode to either New or Append, everything is fine.

Also, what do the Delete/Restore Region buttons do? Clicking them doesn't change anything (peaks or expansion) for me.